19510332-V2.doc 12.07.13 English Operating Manual Heating Immersion Circulator MA JULABO GmbH 77960 Seelbach / Germany

Page 2

Operator responsibility – Safety recommendations 10 2.4. Technical specifications Heating Immersion Circulator MA Working temperature range °C 20

Page 3

MA 11 Safety installations according to IEC 61010-2-010: Excess temperature protection adjustable from 0 °C ... 230 °C Low liquid level protection

Page 4 - Operating manual

Operating instructions 12 Operating instructions 3. Safety notes for the user 3.1. Explanation of safety notes In addition to the safety warnings l

Page 5

MA 13  Do not stay in the area below the unit.  Make sure you read and understand all instructions and safety precautions listed in this manual be

Page 6

Safety notes for the user 14 The user alone is responsible for the handling of these substances! The following questions shall help to recognize possi

Page 7

MA 15 4. Operating controls and functional elements 4.1. Circulator Front view Rear view 1 Mains power switch, illuminated Navigation key

Page 8

Operating controls and functional elements 16 4 1. Key: >Up / Down <temperature – increase/decrease setpoint Push key quickly for sin

Page 9 - 77960 Seelbach / Germany

MA 17 5. Preparations 5.1. Installation Caution: Securely fix the immersion circulator. The heater may not be in contact with the wall of the bath ta

Page 10

Preparations 18 5.2. Bath fluids Caution: Carefully read the safety data sheet of the bath fluid used, particularly with regard to the fire point!

Page 11

MA 19 Caution: Fire or other dangers when using bath fluids that are not recommended: Please contact JULABO before using other than recommended bath

Page 12 - Operating instructions

Operating manual 2 Congratulations! You have made an excellent choice. JULABO thanks you for the trust you have placed in us. This operating manual

Page 13

Preparations 20 5.3.2. Tubing Recommended tubing: Order No. Length Temperature range 8 930 008 1 m CR® tubing 8 mm inner dia. -20 °C to 120

Page 14

MA 21 5.4. Countercooling Notice: Securely attach all tubing to prevent slipping. Observe the laws and regulations of the water distribution company

Page 15

Operating procedures 22 6. Operating procedures 6.1. Power connection Caution:  Only connect the unit to a power socket with earthing contact (PE –

Page 16

MA 23 7. Setting of temperatures The function of the key is configurable. 1. If the key is pressed, normally only one adjustable working tempe

Page 17 - 5. Preparations

Setting of temperatures 24  Refer to page 32 for switch-over to 3-temperature-mode CFG = CONFIGURATION 3SP = 3 SETPOINT Setting of working t

Page 18

MA 25 8. Safety installations, warning functions Check the safety installations at least twice a year! Refer to ( page 14) Settings for the ex

Page 19

Safety installations, warning functions 26 8.1.1. Early warning system, low level protection Warning: For refill always use the same bath fluid typ

Page 20

MA 27 8.2. Switch-over from warning to shutdown function If a shutdown of functional elements (e.g. heater, circulating pump) is required when the

Page 21

Safety installations, warning functions 28 8.3. Over and Sub temperature warning function Factory settings: Over temperature t High 205 °C Sub

Page 22 - 6. Operating procedures

MA 29 9. Menu functions   Menu level 1 The term „Menu functions“ refers to settings such as  >Pu< - Electronically adjustable pump cap

Page 23 - 7. Setting of temperatures

MA 3 TABLE OF CONTENTS Operating manual ...

Page 24

Menu functions 30 9.1. MENU PUMP – Setting of pump pressure Factory setting: stage 1 The capacity of the circulating pump is set by adju

Page 25

MA 31 9.2. MENU Configuration Menu level 1  CFG = MENU CONFIGURATION  A RESET can be effected only in the > < mode. Switch off circula

Page 26

Menu functions 32 9.2.1. Configuration of the mode of the key Factory setting: no Pressing the key normally indicated only one working temper

Page 27

MA 33 Warning: For supervised or unsupervised operation with the “AUTOSTART“ function avoid any hazardous situation to persons or property Take c

Page 28

Menu functions 34 Setting range: 0.1 ... 99.9 Proportional range >Xp< The proportional range is the range below the setpoint in which the

Page 29 - 9. Menu functions

MA 35 9.4. MENU SERIAL - BAUDRATE, PARITY, HANDSHAKE Menu level 1  For communication between circulator and a PC or a superordinated process con

Page 30

Menu functions 36 9.5. MENU ATC - Absolut Temperature Calibration ATC serves to compensate a temperature difference that might occur between circu

Page 31 - key. (refer to page 23)

MA 37  >tYP< - ATC TYPE  The parameter flashes, switch by pressing and  A >1-point<, >2-point< or >3-point< ca

Page 32 -  AUTOSTART off

Menu functions 38 9.5.1. ATC STATUS - YES / NO In the second submenu the ATC function for the temperature sensor selected above is activated >

Page 33

MA 39 9.5.3. Example: 3-point calibration for internal control In the temperature range from 80 °C to 160 °C the calibration curve of the temperat

Page 34

Operating manual 4 8.1.1. Early warning system, low level protection ... 26 8.2. Switch-over from warning

Page 35

Menu functions 40  The value >tt1< is only indicated  Setting >Ct 1< by using the keys.  Integer digits flash, set by pressing

Page 36 - M B (INT) T

MA 41 9.6. MENU LIMITS Menu level 1  >Li< = LIMITS – limitation of operating temperature range Maximum and minimum setpoint Restricti

Page 37

Troubleshooting guide / error messages 42 10. Troubleshooting guide / error messages + Alarm with complete shutdown: If one of the following failu

Page 38

MA 43 Error in A/D converter Excess temperature protector defect. The protection temperature is below the set working temperature setpoint. Set th

Page 39

Electrical connections 44 11. Electrical connections Notice: Use shielded cables only. The shield of the connecting cable is electrically connected

Page 40

MA 45 12. Remote control 12.1. Setup for remote control 9651 RS232C  Check the interface parameters for both interfaces (on circulator and PC) a

Page 41

Remote control 46 12.3. List of commands out commands: Setting parameters or temperature values. Command Parameter Response of circulator version

Page 42

MA 47 Command Parameter Response of circulator in_sp_03 none High temperature warning limit „t High“ in_sp_04 none Low temperature warning limit „

Page 43

Remote control 48 ERROR MESSAGES Description -06 SENSOR DIFFERENCE ALARM Sensor difference alarm. Working temperature and safety sensors report a t

Page 44 - 11. Electrical connections

MA 49 13. JULABO Service – Online remote diagnosis JULABO circulators of the HighTech series are equipped with a black box. This box is implemente

Page 45 -  end of file (; Hex: 0D)

MA 5 Operating manual 1. Intended use JULABO circulators have been designed to control the temperature of specific fluids in a bath tank. JULABO c

Page 46

Cleaning / repairing the unit 50 14. Cleaning / repairing the unit Caution:  Always turn off the unit and disconnect the mains cable from the po

Page 47

Operator responsibility – Safety recommendations 6 2. Operator responsibility – Safety recommendations The products of JULABO ensure safe operation

Page 48

MA 7 Appropriate operation Only qualified personnel is authorized to perform configuration, installation, maintenance and repairs of the circulator. R

Page 49

Operator responsibility – Safety recommendations 8 2.1. Disposal The circulator contains a back-up battery that supplies voltage to memory chips when

Page 50

MA 9 2.2. EC Conformity The products described in the operating instructions conform to the requirements of the following European guidelines: Dire